A lens-coupled scintillation counter in cryogenic environment

arXiv:1101.3254 First posted 2011-01-17 v1 Cited by 2 14 pages, 11 figures
Cite this paper ×
BibTeX
@misc{astoykov2011a,
  title = {A lens-coupled scintillation counter in cryogenic environment},
  author = {A.Stoykov and R.Scheuermann and A.Amato and M.Bartkowiak and J.A.Konter and J.Rodriguez and K.Sedlak},
  year = {2011},
  eprint = {1101.3254},
  archivePrefix = {arXiv},
  primaryClass = {physics.ins-det},
  url = {https://arxiv.org/abs/1101.3254}
}
APA
A.Stoykov, R.Scheuermann, A.Amato, M.Bartkowiak, J.A.Konter, J.Rodriguez, & K.Sedlak (2011). A lens-coupled scintillation counter in cryogenic environment [Preprint]. arXiv. https://doi.org/10.48550/arXiv.1101.3254
GB/T 7714
A.STOYKOV, R.SCHEUERMANN, A.AMATO, et al. A lens-coupled scintillation counter in cryogenic environment[EB/OL]. (2011-01-17)[2026-11-29]. https://arxiv.org/abs/1101.3254.

Abstract

In this work we present an elegant solution for a scintillation counter to be integrated into a cryogenic system. Its distinguishing feature is the absence of a continuous light guide coupling the scintillation and the photodetector parts, operating at cryogenic and room temperatures respectively. The prototype detector consists of a plastic scintillator with glued-in wavelength-shifting fiber located inside a cryostat, a Geiger-mode Avalanche Photodiode (G-APD) outside the cryostat, and a lens system guiding the scintillation light re-emitted by the fiber to the G-APD through optical windows in the cryostat shields. With a 0.8mm diameter multiclad fiber and a 1mm active area G-APD the coupling efficiency of the "lens light guide" is about 50%. A reliable performance of the detector down to 3K is demonstrated.

Post #1 · the paper

Discussion

No comments yet. Start the discussion!