Calculation of thermal parameters of SiGe microbolometers

arXiv:1208.6147 First posted 2012-08-30 v1 Cited by 1 11 pages, 6 figures
Cite this paper ×
BibTeX
@misc{voitsekhovskii2012calculation,
  title = {Calculation of thermal parameters of SiGe microbolometers},
  author = {A. V. Voitsekhovskii and D. V. Grigoryev and V. A. Yuryev and S. N. Nesmelov},
  year = {2012},
  eprint = {1208.6147},
  archivePrefix = {arXiv},
  primaryClass = {cond-mat.mtrl-sci},
  url = {https://arxiv.org/abs/1208.6147}
}
APA
Voitsekhovskii, A. V., Grigoryev, D. V., Yuryev, V. A., & Nesmelov, S. N. (2012). Calculation of thermal parameters of SiGe microbolometers [Preprint]. arXiv. https://doi.org/10.48550/arXiv.1208.6147
GB/T 7714
VOITSEKHOVSKII AV, GRIGORYEV DV, YURYEV VA, et al. Calculation of thermal parameters of SiGe microbolometers[EB/OL]. (2012-08-30)[2026-11-29]. https://arxiv.org/abs/1208.6147.

Abstract

The thermal parameters of a SiGe microbolometer were calculated using numerical modeling. The calculated thermal conduction and thermal response time are in good agreement with the values found experimentally and range between 2x10^-7 and 7x10^-8 W/K and 1.5 and 4.5 ms, respectively. High sensitivity of microbolometer is achieved due to optimization of the thermal response time and thermal conduction by fitting the geometry of supporting heat-removing legs or by selection of a suitable material providing boundary thermal resistance higher than 8x10^-3 cm^2K/W at the SiGe interface.