Performance of an X-ray single pixel TES microcalorimeter under DC and AC biasing

arXiv:1604.02536 First posted 2016-04-09 v1 4 pages, 5 figuers, the Thirteenth International Workshop on Low Temperature Detectors LTD13
Cite this paper ×
BibTeX
@misc{gottardi2016performance,
  title = {Performance of an X-ray single pixel TES microcalorimeter under DC and AC biasing},
  author = {Luciano Gottardi and Jan van der Kuur and Piet A. J. de Korte and Roland Den Hartog and Bob Dirks and Manuela Popescu and Henk F.C. Hoevers and Marcel Bruijn and Maria Parra Borderias and Y. Takei},
  year = {2016},
  eprint = {1604.02536},
  archivePrefix = {arXiv},
  primaryClass = {physics.ins-det},
  url = {https://arxiv.org/abs/1604.02536}
}
APA
Gottardi, L., Kuur, J. V. D., Korte, P. A. J. D., Hartog, R. D., Dirks, B., Popescu, M., Hoevers, H. F., Bruijn, M., Borderias, M. P., & Takei, Y. (2016). Performance of an X-ray single pixel TES microcalorimeter under DC and AC biasing [Preprint]. arXiv. https://doi.org/10.48550/arXiv.1604.02536
GB/T 7714
GOTTARDI L, KUUR JVD, KORTE PAJD, et al. Performance of an X-ray single pixel TES microcalorimeter under DC and AC biasing[EB/OL]. (2016-04-09)[2026-11-29]. https://arxiv.org/abs/1604.02536.

Abstract

We are developing Frequency Domain Multiplexing (FDM) for the read-out of TES imaging microcalorimeter arrays for future X-ray missions like IXO. In the FDMconfiguration the TES is AC voltage biased at a well defined frequencies (between 0.3 to 10MHz) and acts as an AM modulating element. In this paper we will present a full comparison of the performance of a TES microcalorimeter under DC bias and AC bias at a frequency of 370kHz. In both cases we measured the current-to-voltage characteristics, the complex impedance, the noise, the X-ray responsivity, and energy resolution. The behaviour is very similar in both cases, but deviations in performances are observed for detector working points low in the superconducting transition (R/RN < 0.5). The measured energy resolution at 5.89keV is 2.7eV for DC bias and 3.7eV for AC bias, while the baseline resolution is 2.8eV and 3.3eV, respectively.

Post #1 · the paper