High-resolution wide-field OCT angiography with a self-navigation method to correct microsaccades and blinks
@misc{wei2020highresolution,
title = {High-resolution wide-field OCT angiography with a self-navigation method to correct microsaccades and blinks},
author = {Xiang Wei and Tristan T. Hormel and Yukun Guo and Thomas S. Hwang and Yali Jia},
year = {2020},
eprint = {2004.04823},
archivePrefix = {arXiv},
primaryClass = {physics.med-ph},
url = {https://arxiv.org/abs/2004.04823}
}
Wei, X., Hormel, T. T., Guo, Y., Hwang, T. S., & Jia, Y. (2020). High-resolution wide-field OCT angiography with a self-navigation method to correct microsaccades and blinks [Preprint]. arXiv. https://doi.org/10.48550/arXiv.2004.04823
WEI X, HORMEL TT, GUO Y, et al. High-resolution wide-field OCT angiography with a self-navigation method to correct microsaccades and blinks[EB/OL]. (2020-04-09)[2026-11-29]. https://arxiv.org/abs/2004.04823.
Abstract
In this study, we demonstrate a novel self-navigated motion correction method that suppresses eye motion and blinking artifacts on wide-field optical coherence tomographic angiography (OCTA) without requiring any hardware modification. Highly efficient GPU-based, real-time OCTA image acquisition and processing software was developed to detect eye motion artifacts. The algorithm includes an instantaneous motion index that evaluates the strength of motion artifact on en face OCTA images. Areas with suprathreshold motion and eye blinking artifacts are automatically rescanned in real-time. Both healthy eyes and eyes with diabetic retinopathy were imaged, and the self-navigated motion correction performance was demonstrated.