Phase space considerations for a microSAXS beamline located on a diamond Laue side-bounce monochromator

arXiv:2005.12405 First posted 2020-05-26 v1 13 pages, 1 figure
Cite this paper ×
BibTeX
@misc{smilgies2020phase,
  title = {Phase space considerations for a microSAXS beamline located on a diamond Laue side-bounce monochromator},
  author = {Detlef-M. Smilgies},
  year = {2020},
  eprint = {2005.12405},
  archivePrefix = {arXiv},
  primaryClass = {physics.ins-det},
  url = {https://arxiv.org/abs/2005.12405}
}
APA
Smilgies, D.-M. (2020). Phase space considerations for a microSAXS beamline located on a diamond Laue side-bounce monochromator [Preprint]. arXiv. https://doi.org/10.48550/arXiv.2005.12405
GB/T 7714
SMILGIES DM. Phase space considerations for a microSAXS beamline located on a diamond Laue side-bounce monochromator[EB/OL]. (2020-05-26)[2026-11-29]. https://arxiv.org/abs/2005.12405.

Abstract

Flux as well as spatial and angular resolution for a microbeam small-angle x-ray scattering set-up comprising Laue optics and multiple focusing elements are modeled within five-dimensional phase space analysis. A variety of x-ray optics configurations for highest angular resolution and for highest spatial resolution are analyzed.