Evaluation of the KLauS ASIC at low temperature

arXiv:2011.05643 First posted 2020-11-11 v2 15 pages, 12 figures, aticle
Cite this paper ×
BibTeX
@misc{wang2020evaluation,
  title = {Evaluation of the KLauS ASIC at low temperature},
  author = {Wei Wang and Wei Shen and Zhenxiong Yuan and Konrad Briggl and Hans-Christian Schultz-Coulon and Wenqi Yan and Guofu Cao and Zepeng Li and Ming Qi and Liangjian Wen},
  year = {2020},
  eprint = {2011.05643},
  archivePrefix = {arXiv},
  primaryClass = {physics.ins-det},
  url = {https://arxiv.org/abs/2011.05643}
}
APA
Wang, W., Shen, W., Yuan, Z., Briggl, K., Schultz-Coulon, H.-C., Yan, W., Cao, G., Li, Z., Qi, M., & Wen, L. (2020). Evaluation of the KLauS ASIC at low temperature [Preprint]. arXiv. https://doi.org/10.48550/arXiv.2011.05643
GB/T 7714
WANG W, SHEN W, YUAN Z, et al. Evaluation of the KLauS ASIC at low temperature[EB/OL]. (2020-11-11)[2026-11-29]. https://arxiv.org/abs/2011.05643.

Abstract

The Taishan Antineutrino Observatory (TAO) is proposed to first use a cold liquid scintillator detector (-50 ^circC) equipped with large-area silicon photomultipliers (SiPMs) (10 m^2) to precisely measure the reactor antineutrino spectrum with a record energy resolution of < 2% at 1 MeV. The KLauS ASIC shows excellent performance at room temperature and is a potential readout solution for TAO. In this work, we report evaluations of the fifth version of the KLauS ASIC (KLauS5) from room temperature to -50 ^circC with inputs of injected charge or SiPMs. Our results show that KLauS5 has good performance at the tested temperatures with no significant degradation of the charge noise, charge linearity, gain uniformity or recovery time. Meanwhile, we also observe that several key parameters degrade when the chip operates in cold conditions, including the dynamic range and power consumption. However, even with this degradation, a good signal-to-noise ratio and good resolution of a single photoelectron can still be achieved for the tested SiPM with a gain of greater than 1.5x10^6 and even an area of SiPM up to 1 cm^2 in one channel, corresponding to an input capacitance of approximately 5 nF. Thus, we conclude that KLauS5 can fulfill the TAO requirements for precise charge measurement.