A small actively-controlled high-resolution spectrograph based on off-the-shelf components

arXiv:2011.10526 First posted 2020-11-20 v1 21 pages, 9 figures, 2 tables, accepted PASP
Cite this paper ×
BibTeX
@misc{jones2020a,
  title = {A small actively-controlled high-resolution spectrograph based on off-the-shelf components},
  author = {H.R.A. Jones and W. E. Martin and G. Anglada-Escudé and R. Errmann and D. A. Campbell and C. Baker and C. Boonsri and P. Choochalerm},
  year = {2020},
  eprint = {2011.10526},
  archivePrefix = {arXiv},
  primaryClass = {astro-ph.IM},
  url = {https://arxiv.org/abs/2011.10526}
}
APA
Jones, H., Martin, W. E., Anglada-Escudé, G., Errmann, R., Campbell, D. A., Baker, C., Boonsri, C., & Choochalerm, P. (2020). A small actively-controlled high-resolution spectrograph based on off-the-shelf components [Preprint]. arXiv. https://doi.org/10.48550/arXiv.2011.10526
GB/T 7714
JONES H, MARTIN WE, ANGLADA-ESCUDÉ G, et al. A small actively-controlled high-resolution spectrograph based on off-the-shelf components[EB/OL]. (2020-11-20)[2026-11-29]. https://arxiv.org/abs/2011.10526.

Abstract

We present the design and testing of a prototype in-plane echelle spectrograph based on an actively controlled fibre-fed double-pass design. This system aims to be small and efficient with the minimum number of optical surfaces - currently a collimator/camera lens, cross-dispersing prism, grating and a reflector to send light to the detector. It is built from catalogue optical components and has dimensions of approximately 20x30 cm. It works in the optical regime with a resolution of >70,000. The spectrograph is fed by a bifurcated fibre with one fibre to a telescope and the other used to provide simultaneous Thorium Argon light illumination for wavelength calibration. The positions of the arc lines on the detector are processed in real time and commercial auto-guiding software is used to treat the positions of the arc lines as guide stars. The guiding software sends any required adjustments to mechanical piezo-electric actuators which move the mirror sending light to the camera removing any drift in the position of the arc lines. The current configuration using an sCMOS detector provides a precision of 3.5 milli-pixels equivalent to 4 m/s in a standard laboratory environment.

Post #1 · the paper