{"id":"1908.11416","title":"Localization of MEG and EEG Brain Signals by Alternating Projection","authors":["Amir Adler","Mati Wax","Dimitrios Pantazis"],"categories":["eess.SP"],"primary_category":"eess.SP","first_posted":"2019-08-29","updated":"2020-11-25","url":"https://arxiv.vip/abs/1908.11416","arxiv_url":"https://arxiv.org/abs/1908.11416","pdf_url":"https://arxiv.org/pdf/1908.11416","comment_count":0,"abstract":"We present a novel solution to the problem of localization of MEG and EEG brain signals. The solution is sequential and iterative, and is based on minimizing the least-squares (LS)criterion by the Alternating Projection (AP) algorithm, which is well known in the context of array signal processing. Unlike existing scanning solutions belonging to the beamformer and multiple-signal classification (MUSIC) families, the algorithm has good performance in low signal-to-noise ratio (SNR) and can cope with closely spaced sources and any mixture of correlated sources. Results from simulated and experimental MEG data from a real phantom demonstrated robust performance across an extended SNR range, the entire inter-source correlation range, and across multiple sources, with consistently superior localization accuracy than popular scanning methods.","arxiv_comments":null,"journal_ref":null,"doi":null,"license":"http://arxiv.org/licenses/nonexclusive-distrib/1.0/","versions":2,"discussion_url":"https://arxiv.vip/abs/1908.11416#discussion"}