{"id":"2003.09731","title":"SiPM proton irradiation for application in cosmic space","authors":["S. Mianowski","D.M. Borowicz","K. Brylew","A. Dziedzic","M. Grodzicka-Kobylka","A. Korgul","M. Krakowiak","Z. Mianowska","A.G. Molokanov","M. Moszynski","G.V. Mytsin","D. Rybka","K. Shipulin","T. Szczesniak"],"categories":["physics.ins-det","hep-ex"],"primary_category":"physics.ins-det","first_posted":"2020-03-21","updated":"2020-03-21","url":"https://arxiv.vip/abs/2003.09731","arxiv_url":"https://arxiv.org/abs/2003.09731","pdf_url":"https://arxiv.org/pdf/2003.09731","comment_count":0,"abstract":"This paper presents the results of the proton irradiation of silicon photomulipliers (SiPMs) by mono-energetic 170 MeV protons with fluence up to 4.6x10^9 particles/cm^2. In our work, three types of silicon photodetectors from Hamamatsu with areas 3x3 mm^2 and different subpixel sizes of 25x25 mum^2, 50x50 mum^2, and 75x75 mum^2 were used. The changes in the SiPM dark count rate (DCR) spectrum before and after irradiation in temperatures in the range of 20 ^circC to -65 ^circC are presented. The influence of the DCR changes on the energy resolution of the 662 keV gamma line from the ^137Cs for a non-irradiated GAGG:Ce (1%) scintillator is investigated. The time period of usability of the SiPM detector irradiated by protons in cosmic space was estimated.","arxiv_comments":null,"journal_ref":"2020 JINST 15 P03002","doi":"10.1088/1748-0221/15/03/P03002","license":"http://arxiv.org/licenses/nonexclusive-distrib/1.0/","versions":1,"discussion_url":"https://arxiv.vip/abs/2003.09731#discussion"}