Proton radiography to improve proton radiotherapy: Simulation study at different proton beam energies
Proton radiography to improve proton radiotherapy: Simulation study at different proton beam energies
@misc{biegun2016proton,
title = {Proton radiography to improve proton radiotherapy: Simulation study at different proton beam energies},
author = {A.K. Biegun and Jun Takatsu and M-J. van Goethem and E.R. van der Graaf and M. van Beuzekom and J. Visser and S. Brandenburg},
year = {2016},
eprint = {1601.08074},
archivePrefix = {arXiv},
primaryClass = {physics.med-ph},
url = {https://arxiv.org/abs/1601.08074}
}
Biegun, A., Takatsu, J., Goethem, M.-J. V., Graaf, E. V. D., Beuzekom, M. V., Visser, J., & Brandenburg, S. (2016). Proton radiography to improve proton radiotherapy: Simulation study at different proton beam energies [Preprint]. arXiv. https://doi.org/10.48550/arXiv.1601.08074
BIEGUN A, TAKATSU J, GOETHEM MJV, et al. Proton radiography to improve proton radiotherapy: Simulation study at different proton beam energies[EB/OL]. (2016-01-29)[2026-11-29]. https://arxiv.org/abs/1601.08074.
要旨
プロトン治療の質を向上させるためには、X 線コンピュータ断層撮影(CT)画像をプロトンの停止係数マップへの変換精度を高める必要がある。CT 画像から決定されたプロトンの停止係数を用いた患者内でのプロトン射程の計算には、通常 3-4% の系統的な不確実性があり、骨を含む領域では最大 10% にも達する citeUSchneider1995,USchneider1996,WSchneider2000,GCirrone2007,HPaganetti2012,TPlautz2014,GLandry2013,JSchuemann2014。その結果、腫瘍の一部は照射を受けなかったり、健康な組織やリスクのある臓器(例:脳幹)に非常に高い照射量が与えられたりする citeACKnopf2013。高エネルギープロトンの透過放射線写真を用いてプロトンの停止係数を直接測定することは、これらの不確実性を低減し、治療の質を向上させることを可能にする。十分な精度の放射線写真を取得する最良の方法は、フィント(患者)を貫く個々のプロトンの軌跡を追跡することである citeGCirrone2007,TPlautz2014,VSipala2013。我々のシミュレーションでは、フィントの前後に単一プロトンを検出する理想的な位置敏感型検出器を用い、プロトンの残存エネルギーは BaF_2 クリスタルで検出された。透過放射線写真を得るために、3x3 cm^2 の散乱プロトンビームを用いて様々なフィント材料を照射した。シミュレーションは Geant4 シミュレーションパッケージ citeSAgostinelli2003 を用いて行われた。本研究では、プロトン放射線治療で臨床的に利用可能な様々なプロトンビームエネルギーに対するエネルギー損失放射線写真のシミュレーションに焦点を当てている。
To improve the quality of cancer treatment with protons, a translation of X-ray Computed Tomography (CT) images into a map of the proton stopping powers needs to be more accurate. Proton stopping powers determined from CT images have systematic uncertainties in the calculated proton range in a patient of typically 3-4% and even up to 10% in region containing bone citeUSchneider1995,USchneider1996,WSchneider2000,GCirrone2007,HPaganetti2012,TPlautz2014,GLandry2013,JSchuemann2014. As a consequence, part of a tumor may receive no dose, or a very high dose can be delivered in healthy ti\-ssues and organs at risks (e.g. brain stem) citeACKnopf2013. A transmission radiograph of high-energy protons measuring proton stopping powers directly will allow to reduce these uncertainties, and thus improve the quality of treatment. The best way to obtain a sufficiently accurate radiograph is by tracking individual protons traversing the phantom (patient) citeGCirrone2007,TPlautz2014,VSipala2013. In our simulations we have used an ideal position sensitive detectors measuring a single proton before and after a phantom, while the residual energy of a proton was detected by a BaF_2 crystal. To obtain transmission radiographs, diffe\-rent phantom materials have been irradiated with a 3x3 cm^2 scattered proton beam, with various beam energies. The simulations were done using the Geant4 simulation package citeSAgostinelli2003. In this study we focus on the simulations of the energy loss radiographs for various proton beam energies that are clinically available in proton radiotherapy.