薄合金薄膜の定量 X 線蛍光分析における二次蛍光励起の検証
Validation of secondary fluorescence excitation in quantitative X-ray fluorescence analysis of thin alloy films
Validation of secondary fluorescence excitation in quantitative X-ray fluorescence analysis of thin alloy films
@misc{whlisch2020validation,
title = {Validation of secondary fluorescence excitation in quantitative X-ray fluorescence analysis of thin alloy films},
author = {André Wählisch and Cornelia Streeck and Philipp Hönicke and Burkhard Beckhoff},
year = {2020},
eprint = {2003.08930},
archivePrefix = {arXiv},
primaryClass = {cond-mat.mtrl-sci},
url = {https://arxiv.org/abs/2003.08930}
}
Wählisch, A., Streeck, C., Hönicke, P., & Beckhoff, B. (2020). Validation of secondary fluorescence excitation in quantitative X-ray fluorescence analysis of thin alloy films [Preprint]. arXiv. https://doi.org/10.48550/arXiv.2003.08930
WÄHLISCH A, STREECK C, HÖNICKE P, et al. Validation of secondary fluorescence excitation in quantitative X-ray fluorescence analysis of thin alloy films[EB/OL]. (2020-03-19)[2026-11-29]. https://arxiv.org/abs/2003.08930.
要旨
X 線蛍光(XRF)分析は、その非破壊性および容易な自動化により、mum 規模の薄膜の定量分析に広く適用されている。数パーセントの範囲で分析結果の不確実性が低いことが求められる場合、多元素試料における非線形二次蛍光効果は、通常は単純な定量を複雑にする可能性がある。これは、相対寄与が 20% を容易に超えるためである。従来の解決策である、良好な性能を持つ参照試料に依存することは、特に薄膜応用においてその入手可能性が低いという制約に直面している。この課題に対処するため、著者は、顕著な二次蛍光寄与を有する多層合金薄膜の柔軟な製造法を示した。参照なしの XRF 分析を用いて、二次蛍光を包括し、詳細な不確実性評価を含む物理モデルの信頼性を検証した。調査対象の試料は、XRF または他の定量分析用の校正試料として認定された。
X-ray fluorescence (XRF) analysis is a widely applied technique for the quantitative analysis of thin films up to the mum scale because of its non-destructive nature and because it is easily automated. When low uncertainties of the analytical results in the few percent range are required, the non-linear secondary fluorescence effect in multi-elemental samples may complicate an otherwise straightforward quantification, since it can easily exceed a relative contribution of 20%. The conventional solution, to rely on good performing reference samples, is hindered by their low availability, especially for thin film applications. To address this challenge, we demonstrate a flexible production method of multilayered, alloyed thin films with significant secondary fluorescence contributions. We use reference-free XRF analysis to validate the reliability of the physical model for secondary fluorescence, which includes a thorough uncertainty estimation. The investigated specimens are being qualified as calibration samples for XRF or other quantitative analyses.