HARPS のラインプロファイルの計測と特性化:レーザー周波数コンブを用いた
Measuring and characterizing the line profile of HARPS with a laser frequency comb
Measuring and characterizing the line profile of HARPS with a laser frequency comb
@misc{zhao2020measuring,
title = {Measuring and characterizing the line profile of HARPS with a laser frequency comb},
author = {Fei Zhao and G. Lo Curto and L. Pasquini and J. I. González Hernández and J. R. De Medeiros and I. C. Leão and B. L. Canto Martins and R. Rebolo and A. Suárez Mascareño and M. Esposito and A. Manescau and T. Steinmetz and T. Udem and R. Probst and R. Holzwarth and G. Zhao},
year = {2020},
eprint = {2011.03391},
archivePrefix = {arXiv},
primaryClass = {astro-ph.IM},
url = {https://arxiv.org/abs/2011.03391}
}
Zhao, F., Curto, G. L., Pasquini, L., Hernández, J. I. G., Medeiros, J. R. D., Leão, I. C., Martins, B. L. C., Rebolo, R., Mascareño, A. S., Esposito, M., Manescau, A., Steinmetz, T., Udem, T., Probst, R., Holzwarth, R., & Zhao, G. (2020). Measuring and characterizing the line profile of HARPS with a laser frequency comb [Preprint]. arXiv. https://doi.org/10.48550/arXiv.2011.03391
ZHAO F, CURTO GL, PASQUINI L, et al. Measuring and characterizing the line profile of HARPS with a laser frequency comb[EB/OL]. (2020-11-06)[2026-11-29]. https://arxiv.org/abs/2011.03391.
要旨
目的。HARPS(高精度径向速度行星探測器)の 2 次元スペクトル線プロファイルについて研究し、検出器上の位置および線強度の変化に伴うその変動を測定する。線プロファイルおよびその変動の特性解明は、太陽型恒星の可住圏で地球同士の検出に必要な、10^-10 の波長スケール精度または 3.0 cm/s の速度精度の達成に不可欠である。方法。未分解かつ重なりのない線を持つレーザー周波数コム(LFC)を用いて、機器の線プロファイルを検出する。LFC 光(各種中性密度フィルターで減衰させた)を HARPS の対象光ファイバおよび参照光ファイバの両方に注入し、線強度変化に伴う線プロファイルの変動を調べた。線位置、幅、歪み、および分光器および検出器によって誘発された線プロファイルの歪みを測定・特性化するためにモーメント解析を適用した。これに基づき、両ファイバにおけるビームプロファイルの追跡によって点広がり関数の歪みを補正するモデルを確立した。結果。線プロファイルが検出器上の位置および線強度の関数として変動することが示された。これは HARPS 検出器における電荷転送非効率性(CTI)効果と整合する。線位置の推定は線プロファイルに依存しており、したがって線振幅の変化は実測線位置を変化させ、機器の波長スケールの安定性に影響を与える。我々はこれを推定し補正関数を適用して再校正・緩和し、光子ノイズレベルと整合する程度まで低減した。
Aims. We study the 2D spectral line profile of HARPS (High Accuracy Radial Velocity Planet Searcher), measuring its variation with position across the detector and with changing line intensity. The characterization of the line profile and its variations are important for achieving the precision of the wavelength scales of 10^-10 or 3.0 cm/s necessary to detect Earth-twins in the habitable zone around solar-like stars. Methods. We used a laser frequency comb (LFC) with unresolved and unblended lines to probe the instrument line profile. We injected the LFC light (attenuated by various neutral density filters) into both the object and the reference fibres of HARPS, and we studied the variations of the line profiles with the line intensities. We applied moment analysis to measure the line positions, widths, and skewness as well as to characterize the line profile distortions induced by the spectrograph and detectors. Based on this, we established a model to correct for point spread function distortions by tracking the beam profiles in both fibres. Results. We demonstrate that the line profile varies with the position on the detector and as a function of line intensities. This is consistent with a charge transfer inefficiency (CTI) effect on the HARPS detector. The estimate of the line position depends critically on the line profile, and therefore a change in the line amplitude effectively changes the measured position of the lines, affecting the stability of the wavelength scale of the instrument. We deduce and apply the correcting functions to re-calibrate and mitigate this effect, reducing it to a level consistent with photon noise.