양성자 영상법으로 양성자 방사선 치료의 개선: 다른 양성자 빔 에너지에서의 시뮬레이션 연구
Proton radiography to improve proton radiotherapy: Simulation study at different proton beam energies
Proton radiography to improve proton radiotherapy: Simulation study at different proton beam energies
@misc{biegun2016proton,
title = {Proton radiography to improve proton radiotherapy: Simulation study at different proton beam energies},
author = {A.K. Biegun and Jun Takatsu and M-J. van Goethem and E.R. van der Graaf and M. van Beuzekom and J. Visser and S. Brandenburg},
year = {2016},
eprint = {1601.08074},
archivePrefix = {arXiv},
primaryClass = {physics.med-ph},
url = {https://arxiv.org/abs/1601.08074}
}
Biegun, A., Takatsu, J., Goethem, M.-J. V., Graaf, E. V. D., Beuzekom, M. V., Visser, J., & Brandenburg, S. (2016). Proton radiography to improve proton radiotherapy: Simulation study at different proton beam energies [Preprint]. arXiv. https://doi.org/10.48550/arXiv.1601.08074
BIEGUN A, TAKATSU J, GOETHEM MJV, et al. Proton radiography to improve proton radiotherapy: Simulation study at different proton beam energies[EB/OL]. (2016-01-29)[2026-11-29]. https://arxiv.org/abs/1601.08074.
초록
양성자 치료의 질을 향상시키기 위해 X 선 컴퓨터 단층 촬영 (CT) 이미지를 양성자 정지력 지도로 변환하는 정확도가 더 높아져야 한다. CT 이미지로부터 결정된 양성자 정지력은 환자 내 계산된 양성자 경로에 대해 일반적으로 3-4% 의 체계적 불확실성을 가지며, 뼈가 포함된 영역에서는 최대 10% 까지 도달할 수 있다 citeUSchneider1995,USchneider1996,WSchneider2000,GCirrone2007,HPaganetti2012,TPlautz2014,GLandry2013,JSchuemann2014. 그 결과 종양의 일부는 투여량을 받지 않거나, 건강한 조직 및 위험 장기 (예: 뇌간) 에 매우 높은 투여량이 전달될 수 있다 citeACKnopf2013. 고에너지 양성자의 투과 방사선 사진은 양성자 정지력을 직접 측정하여 이러한 불확실성을 줄이고, 결과적으로 치료의 질을 향상시킬 수 있다. 충분히 정확한 투과 방사선 사진을 얻는 가장 좋은 방법은 환자를 통과하는 개별 양성자를 추적하는 것이다 citeGCirrone2007,TPlautz2014,VSipala2013. 우리의 시뮬레이션에서는 환자를 통과하기 전과 후에 단일 양성자를 측정하는 이상적인 위치 감지 검출器和 양성자의 잔여 에너지를 BaF_2 결정체로 검출하는 방식을 사용하였다. 투과 방사선 사진을 얻기 위해 다양한 phantom 재료를 3x3 cm^2 산란 양성자 빔으로 조사하였다. 시뮬레이션은 Geant4 시뮬레이션 패키지를 사용하여 수행되었다 citeSAgostinelli2003. 본 연구에서는 양성자 방사선 치료에서 임상적으로 이용 가능한 다양한 양성자 빔 에너지에 대한 에너지 손실 방사선 사진의 시뮬레이션에 초점을 맞추었다.
To improve the quality of cancer treatment with protons, a translation of X-ray Computed Tomography (CT) images into a map of the proton stopping powers needs to be more accurate. Proton stopping powers determined from CT images have systematic uncertainties in the calculated proton range in a patient of typically 3-4% and even up to 10% in region containing bone citeUSchneider1995,USchneider1996,WSchneider2000,GCirrone2007,HPaganetti2012,TPlautz2014,GLandry2013,JSchuemann2014. As a consequence, part of a tumor may receive no dose, or a very high dose can be delivered in healthy ti\-ssues and organs at risks (e.g. brain stem) citeACKnopf2013. A transmission radiograph of high-energy protons measuring proton stopping powers directly will allow to reduce these uncertainties, and thus improve the quality of treatment. The best way to obtain a sufficiently accurate radiograph is by tracking individual protons traversing the phantom (patient) citeGCirrone2007,TPlautz2014,VSipala2013. In our simulations we have used an ideal position sensitive detectors measuring a single proton before and after a phantom, while the residual energy of a proton was detected by a BaF_2 crystal. To obtain transmission radiographs, diffe\-rent phantom materials have been irradiated with a 3x3 cm^2 scattered proton beam, with various beam energies. The simulations were done using the Geant4 simulation package citeSAgostinelli2003. In this study we focus on the simulations of the energy loss radiographs for various proton beam energies that are clinically available in proton radiotherapy.