质子成像术改善质子放疗:不同质子束能量下的模拟研究
Proton radiography to improve proton radiotherapy: Simulation study at different proton beam energies
Proton radiography to improve proton radiotherapy: Simulation study at different proton beam energies
@misc{biegun2016proton,
title = {Proton radiography to improve proton radiotherapy: Simulation study at different proton beam energies},
author = {A.K. Biegun and Jun Takatsu and M-J. van Goethem and E.R. van der Graaf and M. van Beuzekom and J. Visser and S. Brandenburg},
year = {2016},
eprint = {1601.08074},
archivePrefix = {arXiv},
primaryClass = {physics.med-ph},
url = {https://arxiv.org/abs/1601.08074}
}
Biegun, A., Takatsu, J., Goethem, M.-J. V., Graaf, E. V. D., Beuzekom, M. V., Visser, J., & Brandenburg, S. (2016). Proton radiography to improve proton radiotherapy: Simulation study at different proton beam energies [Preprint]. arXiv. https://doi.org/10.48550/arXiv.1601.08074
BIEGUN A, TAKATSU J, GOETHEM MJV, et al. Proton radiography to improve proton radiotherapy: Simulation study at different proton beam energies[EB/OL]. (2016-01-29)[2026-11-29]. https://arxiv.org/abs/1601.08074.
摘要
为提升质子治疗的质量,需将 X 射线计算机断层扫描(CT)图像更准确地转换为质子阻止本领图。由 CT 图像确定的质子阻止本领在计算患者体内的质子射程时存在系统不确定性,通常为 3%–4%,在含骨区域甚至高达 10% citeUSchneider1995,USchneider1996,WSchneider2000,GCirrone2007,HPaganetti2012,TPlautz2014,GLandry2013,JSchuemann2014。因此,肿瘤的一部分可能未接受剂量,或健康组织和危及器官(如脑干) citeACKnopf2013 可能接受极高剂量。直接测量质子阻止本领的高能质子透射 X 射线成像可减少此类不确定性,从而改善治疗效果。获取足够准确透射成像的最佳方法是追踪穿过体模(患者)的单个质子 citeGCirrone2007,TPlautz2014,VSipala2013。在我们的模拟中,使用了理想的位置敏感探测器在体模前后测量单个质子,而质子剩余能量由 BaF_2 晶体探测。为获得透射成像,使用 3x3 cm^2 散射质子束对多种体模材料进行辐照,并采用不同的束流能量。模拟使用 Geant4 模拟包 citeSAgostinelli2003 完成。本研究聚焦于模拟质子放疗中临床可用的各种质子束能量的能量损失成像。
To improve the quality of cancer treatment with protons, a translation of X-ray Computed Tomography (CT) images into a map of the proton stopping powers needs to be more accurate. Proton stopping powers determined from CT images have systematic uncertainties in the calculated proton range in a patient of typically 3-4% and even up to 10% in region containing bone citeUSchneider1995,USchneider1996,WSchneider2000,GCirrone2007,HPaganetti2012,TPlautz2014,GLandry2013,JSchuemann2014. As a consequence, part of a tumor may receive no dose, or a very high dose can be delivered in healthy ti\-ssues and organs at risks (e.g. brain stem) citeACKnopf2013. A transmission radiograph of high-energy protons measuring proton stopping powers directly will allow to reduce these uncertainties, and thus improve the quality of treatment. The best way to obtain a sufficiently accurate radiograph is by tracking individual protons traversing the phantom (patient) citeGCirrone2007,TPlautz2014,VSipala2013. In our simulations we have used an ideal position sensitive detectors measuring a single proton before and after a phantom, while the residual energy of a proton was detected by a BaF_2 crystal. To obtain transmission radiographs, diffe\-rent phantom materials have been irradiated with a 3x3 cm^2 scattered proton beam, with various beam energies. The simulations were done using the Geant4 simulation package citeSAgostinelli2003. In this study we focus on the simulations of the energy loss radiographs for various proton beam energies that are clinically available in proton radiotherapy.